ETS-Lindgren FM5004 Field Monitor Firmware 5.4

Manufacturer:

Description

DOWNLOAD NOW

Improvements :

1) Clipped battery percentage values to 100
2) Automatically pick up a probe after a reset from the front panel.
3) Prevent loss of communication with the HI-6005 after a temperature request from the front panel.
4) Cleared OVER RANGE not getting cleared on the averaging screen.
5) Resolved issues with the RDIS command
6) Timeouts on the probe cause timeouts in the FM5004. If these timeouts are not detected at the GPIB level, this can lead to remnants of readings being left in the output GPIB buffer. This causes confusion and it happens several times, can lead to what appears as a lockup.
7) Auto Ranging was improved when the probe was in high power fields

Note:

The PC driver application must be written properly so that the GPIB RDIS,0 command is used to get CXYZ values from the monitor. After receiving a RDIS,0 command, the monitor sends CXYZ values in 4 different lines. The driver must use 4 different reads in order to get each line properly. If a timeout occurs in between any of the reads, the next read will get information that was meant to the last read. For instance, the second read should get an X value. If there is an unhandled timeout, the next read, instead of getting a Y value will get an X value. This problem can lead to out-of-sync RDIS,0 responses where the next RDIS,0 command will not start with a C value, but with the Z value from previous RDIS,0 commands. If enough timeouts occurs, the leftovers can cause the monitor to hang.

About Device Firmware:

Applying a newer firmware version than the one already installed on your unit can bring various enhancements, include workarounds for diverse problems encountered by the device, and improve or add newly developed features.

On the other hand, downgrading the firmware version can recover the device’s functionality in the event of a software update either turning up faulty or causing the unit’s overall performance to drop. However, bear in mind that applying an older firmware isn’t recommended and might not always be possible.

When it comes to the update steps, due to the large variety of devices and numerous ways for installing a new firmware, it is recommended that you carefully read and understand the upgrade steps, and consider applying a different firmware only when you have familiarized yourself with the update method.

As the installation process is most of the times quite risky, this step should be taken by users that have the ability to successfully complete the update; regular users may initiate it at their own risk. Moreover, it’s best that this task be performed in a steady power environment such as the one ensured by a UPS unit.

Therefore, if you consider applying this release, hit the download button, take into account all the aforementioned aspects, and update the device’s firmware. Also make sure you constantly check with our website to ensure that you don’t miss a single new release.

ETS-Lindgren FM5004 Firmware FM5004 Firmware Update FM5004 Field Monitor Firmware FM5004 Field Monitor Firmware ETS-Lindgren

  CATEGORY:
Firmware
  COMPATIBLE WITH:
OS Independent
  file size:
158 KB
  filename:
rev_5_4_firmwareAndNotes.zip